Specifications of hood BI-3000 BI-2000 BI-1000 BI-800
Mattress to Hood 450 mm 450 mm 450 mm 410 mm
Observation side of hood Two incline Two incline Two incline Two incline
Operate port 6 (2 Iris ports) 6 (2 Iris ports) 6 (2 Iris ports) 4 (2 Iris ports)
Tubing grommets 4 4 4 2
Front access panel 826×310 mm 826×310 mm 826×310 mm 826×280 mm
Mattress Tray Size 630×360 mm 630×360 mm 630×360 mm 630×360 mm
Technical Specification
Power requirement 400VA 400VA 400VA 400VA
Temperature control mode Air and skin servo control Air and skin servo control Air servo control Air servo control
AirTemp Control Range 20°C~37°C 20°C~37°C 20°C~37°C 20°C~37°C
Temperature override mode to 20°C~39°C 20°C~39°C 20°C~39°C ----
SkinTemp Control Range 30°C~37°C 30°C~37°C ---- ----
Temperature override mode to 30°C~38°C 30°C~38°C ---- ----
Skin temp sensor accuracy 0.3°C 0.3°C ---- ----
Humidity display range 5%~99% ---- ---- ----
Temperature fluctuation 0.5°C 0.5°C 0.5°C 0.5°C
Mattress temperature uniformity 0.8°C 0.8°C 0.8°C 0.8°C
Temperature rise time 30 min 30 min 30 min 30 min
Noise level within hood environment 55dB 55dB 55dB 55dB
Alarm
Power failure alarm Yes Yes Yes Yes
Fan failure alarm Yes Yes Yes Yes
Sensor failure alarm Yes Yes Yes Yes
High air temperature alarm +1.5°C +1.5°C +1.5°C +1.5°C
Low air temperature alarm -2.5°C -2.5°C -2.5°C -2.5°C
High skin temperature alarm +1.0°C +1.0°C ---- ----
Low skin temperature alarm -1.0°C -1.0°C ---- ----
Over air temperature alarm 38.0°C; 40.0°C 38.0°C; 40.0°C 38.0°C; 40.0°C 38.0°C
Over skin temperature alarm 38.0°C; 39.0°C 38.0°C; 39.0°C 38.0°C; 39.0°C ----
System failure alarm Yes Yes Yes ----
Alarm silence function Yes Yes Yes ----
Operational condition
Environment temperature +20°C~+30°C +20°C~+30°C +20°C~+30°C +20°C~+30°C
Environment relative humidity 5%~99% 5%~99% 5%~99% 5%~99%
Environment air velocity of flow <0.3m/s <0.3m/s <0.3m/s <0.3m/s
Others
RS-232Connector Yes Yes Yes -----
Rail Yes Yes Yes -----
Infant bed tilt adjusting ±10° step-less ±10° step-less ±10° step-less -----
Mattress tray Max. load 10kg 10kg 10kg 10kg
Transfusion pole Max. load 2kg 2kg 2kg 2kg
Cabinet Max. load 6kg 6kg 6kg ----